1. Measurement is based on the difference in thermal conductivity between hydrogen and other gases;
2. Uses our independently developed high-stability glass-coated platinum (Pt) resistance sensor for high-precision, low-cost measurement;
3. The platinum resistor offers excellent durability and requires almost no maintenance;
4. The sensor is separate from the controller for easy installation and operation;
5. The sensor housing is made of SUS316L; it is compact and its fittings are customizable, making it suitable for various narrow and enclosed environments;
6. Excellent measurement accuracy: ±1% F.S. (full scale);
7. Fast response: 90% response time of less than 10 seconds.
1. Monitoring and control of hydrogen concentration in various gas mixtures;
2. Hydrogen concentration monitoring in nitriding furnaces and control of the ammonia feed rate;
3. Purity and concentration monitoring for fuel cell reformers;
4. Purity and concentration monitoring of by-product hydrogen (chemical plants, steel mills, etc.);
5. Hydrogen concentration monitoring in semiconductor equipment;
6. Hydrogen concentration monitoring in ammonia, methane and methanol synthesis processes;
7. Concentration monitoring in continuous hydrogen furnaces.
| Item | Parameter |
| Measurement principle | Thermal conductivity |
| Gas types analyzed | Hydrogen-containing gas mixtures (free of oxygen, O₂) |
| Gas sampling method | Natural diffusion |
| Measuring range | 0-100% (other ranges available on request) |
| Resolution | 0.1% |
| Accuracy | ±1% F.S. |
| Sample gas temperature | Below 90 ℃; interfering gases: CO₂ (above 5%) / CH₄ (above 10%) / water vapor (above 5 kPa) - compensation correction table provided |
| Maximum sample gas pressure | 300 kPa |
| Materials of gas-contacting parts | SUS316 stainless steel* / glass* / fluororubber |
| Output signal | RS485 / 4-20 mA (permissible load resistance 500 Ω) |
| Power supply | DC 24V / 10W max |
| Warm-up time | 30 min |
| Ambient temperature and humidity | 5-45 ℃, below 90% RH |
| Installation method | Threaded mounting of the sensing unit (standard: NPT 1/4-inch male thread, customizable) |
| Dimensions (controller) | 151 × 83 × 51 mm(H×W×D) |
| Dimensions (sensor) | 105 × 40 × 40 mm(H×W×D) |
| Controller housing | Aluminum alloy enclosure (indoor type) |